3-2
3.1 Unpacking and Checking
Before unpacking, examine the packing case carefully for signs of damage. If any damage is
detected, contact the carrier or our company.
If the packing case is intact, open the package and remove the instrument and accessories
carefully. Check all materials against the packing list and check for any mechanical damage.
Contact our Customer Service Department in case of any problem.
NOTE
z Please save the packing case and packaging material for future transport and
storage.
WARNING
z Be sure to keep the packaging materials from children’s reach.
z Disposal of the packaging materials shall comply with your local requirements.
z The equipment might be contaminated in storage, transport or when used. Verify
the package and the single use accessories are intact. In case of any damage, do not
apply it to patients.
3.2 Environmental Requirements
The operating environment of the monitor must meet the requirements specified in this manual.
The environment where this monitor is to be used should be free from noise, vibration, dust, and
corrosive or explosive and inflammable substances. For a cabinet mounted installation, allow
sufficient room at the front and the rear of the cabinet for operation, maintenance and servicing.
Besides, allow at least 2 inches clearance around the instrument for proper air circulation.
Condensation can form when the monitor is moved from one location to another, and being
exposed to differences in humidity or temperature. Make sure that during operation the
instrument is free from condensation.