15-16
15.3.4 CO
2
User Maintain Menu
Select CO2 USER MAINTAIN >> in USER MAINTAIN menu. The following menu appears.
Figure 15-10 CO
2
User Maintain Menu
The following information is displayed in the menu above.
CO2 The currently measured CO2 content.
Unit: %.
In the field at the right of the CO
2
value, you can select a standard gas containing a certain
amount of CO
2
for calibration.
BAROMETRIC The currently measured barometric pressure.
Unit: mmHg.
CONFIRM CAL This option allows you to confirm the calibration of the CO2
module.