Mindray DPM 1 Instrukcja Operatora Strona 124

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 274
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 123
10-20
10.5.3 Adjusting ST Measurement Point
The ST measurement for each beat complex is the ventical difference between two measurement
points. The isoelectric point provides the baseline for the measurement and the ST point provides
the other measurement point.
Figure 10-14 ST Measurement Point
The ST measurement points need to be adjusted when you start monitoring, and if the patient’s
heart rate or ECG morphology changes siginificantly. Artifactual ST segment depression or
elevation may occur if the ISO or the ST point is incorrectly set.
NOTE
z Abnormal QRS complex is not considered in ST analysis.
R wave
P wave
T wave
Base point ISO
ST measurement point
ST measurement value
Przeglądanie stron 123
1 2 ... 119 120 121 122 123 124 125 126 127 128 129 ... 273 274

Komentarze do niniejszej Instrukcji

Brak uwag