A-17
A.15 AG Specifications
Measurement technique Infrared absorption
Measurement mode Sidestream
AG functions
Meets requirements of ISO9918, ISO11196, EN12598 and
ISO7767
Warm-up time
45 seconds (warming-up status)
10 minutes (ready-to-measure status)
Adult/Pediatric 120, 150, 200 ml/minute (user-selectable)
Sampling flow
(sidestream)
Neonatal 70, 90, 120 ml/minute (user-selectable)
Gas type CO2, N2O, O2 (optional), Des, Iso, Enf, Sev, Hal
Measurement range
CO2:
N2O:
Des:
Sev:
Enf, Iso, Hal:
O2:
AwRR:
0 to 30%
0 to 100%
0 to 30%
0 to 30%
0 to 30%
0 to 100%
2 to 100 BrPM
Resolution
CO2:
AwRR:
1 mmHg
1 BrPM
Gas Range (%REL) Precision (%ABS)
0 to 1
±0.1
1 to 5
±0.2
5 to 7
±0.3
7 to 10
±0.5
CO2
> 10 Not specified
0 to 20
±2
N2O
20 to 100
±3
0 to 1
±0.15
1to 5
±0.2
5 to 10
±0.4
10 to 15
±0.6
15 to 18
±1
Des
>18 Not specified
0 to 1
±0.15
1 to 5
±0.2
Precision
Sev
5 to 8
±0.4