Mindray DPM 1 Instrukcja Operatora Strona 168

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 274
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 167
12-10
Figure 12-5 Replacing the Rubber Bladder
To replace the rubber bladder in the cuff:
1. Place the bladder on top of the cuff so the rubber tubes line up with the large opening on the
long side of the cuff.
2. Roll the bladder lengthwise and insert it into the opening on the long side of the cuff.
3. Hold the tubes and the cuff and shake the complete cuff until the bladder is in position.
4. Thread the rubber tubes from inside the cuff, and out through the small hole under the
internal flap.
CAUTION
z Some disinfectants may cause skin irritation. Please rinse cuff thoroughly with
water to remove any residual disinfectants. Using dark colored disinfectants
may stain the cuffs. Test a single cuff to ensure that no damage will occur.
Disposable Blood Pressure Cuffs
Disposable cuffs are intended for single patient use only. Do not sterilize or use autoclave
sterilization for disposable cuffs.
NOTE
z Disposable blood pressure cuffs must be recycled or disposed of properly.
Przeglądanie stron 167
1 2 ... 163 164 165 166 167 168 169 170 171 172 173 ... 273 274

Komentarze do niniejszej Instrukcji

Brak uwag