C-2
C.2.1 General Alarm Messages of Parameter Modules
Alarm message A B L Cause Action
XX INIT ERR N Yes No H An error occurs to
the XX module
during initialization.
Note: N stands for error code.
XX COMM STOP
No No H Failure in
communication
between XX module
and the main board.
XX COMM
ERROR
No No H Failure in
communication
between XX module
and the main board.
Restart the monitor. If
the problem persists,
contact our service
personnel for repair.
XX ALM LMT
ERR
No No H The alarm limit of
the XX parameter is
changed
inadvertently.
XX EXCEED No No H The measured XX
value exceeds the
measurement range.
If the problem
persists, contact our
service personnel for
repair.
C.2.2 NIBP Module Alarm Messages
Alarm message A B L Cause Action
NIBP SELFTEST
ERR
Yes Yes H An error occurs
during NIBP module
initialization.
Select NIBP RESET
in the MAINTAIN
menu. If the problem
still exists, contact
our service personnel
for repair.
NIBP INIT ERR Yes Yes H An error occurs
during NIBP module
initialization.
Restart the monitor. If
the problem still
exists, contact our
service personnel for
repair.