Alarms Operations
2 - 74 0070-10-0666-01 Trio™ Operating Instructions
TEMP ALARM SETUP Menu Selections
FIGURE 2-54 TEMP Alarm Setup Menu
ALM SEL
(Alarm Selection)
Choose TEMP ALARM SETUP.
ALM Allows the user to turn the alarm ON or OFF. Choose ON to
enable the alarm; choose OFF to disable the alarm. If the alarm is
set to OFF, the alarm OFF symbol will display within the
parameter tile. The alarm is activated when the measurement
exceeds the high or low limits.
ALM PRIORITY Allows the user to select the priority of the alarm. Choices are 1,
2, and 3. Priority 1 alarms are considered the most serious.
ALM PRINT Allows the user to enable or disable automatic printing during a
physiological alarm condition. Choose ON to enable printing
upon alarm. Choose OFF to disable printing upon alarm. See
“Recorder (Optional)” on page 2-82 for details.
TEMP ALM HI Allows the user to set the upper limit of the TEMP alarm. (See
“Temperature Alarm Limits” on page 2-54.)
TEMP ALM LO Allows the user to set the lower limit of the TEMP alarm. (See
“Temperature Alarm Limits” on page 2-54.)